Title :
ANA Calibration Method for Measurements of Dielectric Properties
Author :
Kraszewski, Andrzej ; Stuchly, Maria A. ; Stuchly, Stanislaw S.
fDate :
6/1/1983 12:00:00 AM
Abstract :
Routine network analyzer calibration procedures in measurements of the dielectric properties of materials using an open-ended coaxial line probe are frequently inadequate and limit the accuracy of measurements. A calibration method, which makes use of liquids whose properties are well known, is proposed to alleviate this limitation. It is shown that even one liquid used as a standard in place of a matched load greatly improves the accuracy of measurements. Theoretical relationships and experimental results as well as some practical suggestions related to the application of this method are given.
Keywords :
Calibration; Coaxial components; Dielectric measurements; Frequency; Impedance measurement; Liquids; Measurement standards; Permittivity measurement; Probes; Reflection;
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
DOI :