• DocumentCode
    1204123
  • Title

    Computer Detection of Distortion in Arterial Pressure Signals

  • Author

    Klee, George ; Ackerman, Eugene ; Leonard, Arnold

  • Author_Institution
    Division of Health Computer Sciences, Medical School, University of Minnesota, Minneapolis, Minn. 55455.
  • Issue
    1
  • fYear
    1974
  • Firstpage
    73
  • Lastpage
    75
  • Abstract
    An application of the mathematical concept of curvature to the detection of distortion in arterial pressure signals is described. Curvature is defined and examples of digitally filtered pressure tracings show the relationship between curvature and the degree of damping. Even though curvature is also a function of the cardiovascular system producing the signal, the average sum of curvature can generally distinguish control signals from signals distorted by moderate amounts of blood or air in the catheter systems. Selected pressure signals with their associated sums of curvature show the properties and limitations of curvature for discriminating between the changes in the cardio-vascular system and distortion of pressure signals.
  • Keywords
    Biomedical measurements; Blood pressure; Cardiology; Cardiovascular system; Catheters; Computerized monitoring; Damping; Distortion; Pressure control; Pressure measurement; Animals; Blood Pressure Determination; Cardiovascular Physiology; Computers; Dogs; Humans; Monitoring, Physiologic;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.1974.324372
  • Filename
    4120728