• DocumentCode
    1204132
  • Title

    Mechanical Damping of the Manometric System Used in the Pressure Gradient Technique

  • Author

    Lapointe, A.C. ; Roberge, F.A.

  • Author_Institution
    Department of Physiology, Faculty of Medicine, Ecole Polytechnique, University of Montreal, Montreal, P. Q., Canada.
  • Issue
    1
  • fYear
    1974
  • Firstpage
    76
  • Lastpage
    78
  • Abstract
    A simple method for introducing mechanical damping in catheter-manometer systems so that a flat frequency response throughout 30 Hz can be realized easily is described. This allows reliable computation of instantaneous blood velocity by the pressure gradient technique, since the bandwidth available, dc to 22 Hz, allows accurate recording of the pressure gradient, even when long catheters are used. Similarly, the first time derivative of the intraventricular pressure pulse can be obtained without difficulty.
  • Keywords
    Bandwidth; Biomedical engineering; Blood; Catheters; Damping; Frequency response; Humans; Needles; Transducers; Valves; Animals; Blood Flow Velocity; Blood Pressure Determination; Dogs; Heart Catheterization; Heart Ventricles; Manometry; Pulse; Transducers;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.1974.324373
  • Filename
    4120729