DocumentCode
1204152
Title
Experimental Study of the Frequency Stability of a Maser Oscillator Operated with an External Feedback Loop
Author
Tetu, Michel ; Tremblay, Pierre ; Lesage, Paul ; Petit, Pierre
Volume
32
Issue
3
fYear
1983
Firstpage
410
Lastpage
413
Abstract
In this paper, we report the results of an experimental study of the short-and mid-term frequency stability for both a hydrogen maser and a rubidium maser operated with an external feedback loop to modify the cavity quality factor. A revised version of the theoretical expression of the frequency stability for this type of maser is given and a numerical solution for various maser parameters is calculated. The predicted frequency stability exhibits an optimum when the cavity Q is varied. The experimental results presented in this study agree with this conclusion.
Keywords
Feedback loop; Frequency; Hydrogen; Masers; Microwave amplifiers; Oscillators; Q factor; Space technology; Stability; Temperature;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/TIM.1983.4315096
Filename
4315096
Link To Document