Abstract
Fixed frequency, point-by-point calibration is outlined for a reference standard directional coupler to be used in steppedfrequency measurement of 40-dB, or greater, coupling factors of directional couplers. If a reference directional coupler has a 40-dB coupling factor, a main-line waveguide output, and a sexless coaxial side-arm output, the best accuracy obtainable in measurement of its coupling factor is estimated to be ±0.03 dB.
Keywords
Attenuation measurement; Calibration; Coaxial components; Detectors; Directional couplers; Equations; Frequency measurement; Impedance; Measurement standards; Microwave measurements;