DocumentCode
1204436
Title
A Simple Stray-Free Capacitance Meter by Using an Operational Amplifier
Author
Kuroda, Seiji
Volume
32
Issue
4
fYear
1983
Firstpage
512
Lastpage
513
Abstract
A simple operational amplifier circuit known as a charge amplifier has an advantage of measuring capacitance independent of stray capacitance. Capacitance down to 0.03 pF could be experimentally measured with ease and the theoretical limit of measurement is discussed. Also the direct measurement of the electrostatic induction coefficients of a system of conductors by the circuit is introduced.
Keywords
Bridge circuits; Cable shielding; Capacitance measurement; Capacitors; Current measurement; Cutoff frequency; Electrostatic measurements; Operational amplifiers; Voltage; Wires;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/TIM.1983.4315128