Title :
EEG Spectra Measured with a Distorting System
Author_Institution :
Department of Clinical Neurophysiology, Sahlgren Hospital, Gothenburg, Sweden.
Abstract :
The distortion effects upon the measurement of EEG spectra are investigated. A non-linear model of zero-memory type is used for the input-output relation of the measurement system. The model is identified by means of measured second and third harmonics. Examples illustrate the distortion effects upon power spectra of waking infant and adult EEG´s.
Keywords :
Batteries; Biomedical measurements; Brain modeling; Catheters; Distortion measurement; Electroencephalography; Heart; Power harmonic filters; Power supplies; Power system modeling; Adult; Electroencephalography; Humans; Infant; Infant, Newborn; Models, Biological;
Journal_Title :
Biomedical Engineering, IEEE Transactions on
DOI :
10.1109/TBME.1975.324545