• DocumentCode
    1206169
  • Title

    Precise Voltage to Frequency Converter for Telemetry Applications of Strain Gage Pressure Transducers

  • Author

    Grover, Thomas P.

  • Author_Institution
    Utah Biomedical Test Laboratory, University of Utah Research Institute, Salt Lake City, Utah 84108.
  • Issue
    5
  • fYear
    1975
  • Firstpage
    441
  • Lastpage
    443
  • Abstract
    A voltage to frequency converter suitable for telemetric applications of strain gage pressure transducers is described and analyzed. The circuit uses a voltage reference IC and a temperature dependent op amp bias current to achieve zero pressure offsets less than 3 torr over a 25°C temperature range and a 1 volt supply voltage range.
  • Keywords
    Bridge circuits; Capacitive sensors; Frequency conversion; Operational amplifiers; Pulse circuits; Pulse modulation; Switches; Telemetry; Transducers; Voltage; Electricity; Pressure; Telemetry; Temperature; Transducers;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.1975.324521
  • Filename