• DocumentCode
    1206365
  • Title

    A Spectrophotometer for the Direct Determination of Sample Extinction by Simultaneous Measurement at Multiple Wavelengths and Application of Allen´s Correction

  • Author

    Potter, James B. ; Brown, James B. ; Patrick, James F. ; Schramp, Joseph M.

  • Author_Institution
    Department of Electrical Engineering, University of Melbourne, Melbourne, Australia.
  • Issue
    6
  • fYear
    1975
  • Firstpage
    528
  • Lastpage
    532
  • Abstract
    A spectrophotometer for the direct determination of sample extinction by simultaneous measurement at three wavelengths and application of Allen´s correction has been developed. White light is passed, first through a blue heat filter, and then through the sample and finally through a trichromator which selects the required three wavelengths. These wavelengths illuminate the base region of three FPT 100 type photo diodes which operate as linear transducers converting the incident light intensity to a current which is converted to a voltage in head amplifiers. A computation sub-unit together with a logarithmic amplifier generate the function required for the Allen correction, and the corrected extinction value is displayed digitally on neon display tubes.
  • Keywords
    Absorption; Computer displays; Diodes; Filters; Impurities; Pregnancy; Transducers; Voltage; Wavelength conversion; Wavelength measurement; Colorimetry; Estrogens; Models, Theoretical; Spectrophotometry;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.1975.324476
  • Filename
    4120983