DocumentCode
1206630
Title
Minimizing Silver-Silver Chloride Electrode Impedance
Author
Getzel, Walter A. ; Webster, John G.
Author_Institution
Department of Biomedical Engineering, Marquette University, Milwaukee, WI 53233.
Issue
1
fYear
1976
Firstpage
87
Lastpage
88
Abstract
We sought to minimize silver-silver chloride electrode impedance by chemical treatment. Chloriding pure silver electrodes lowered the impedance, as noted by Geddes [1]. Immersing the chlorided electrodes in photographic developer lowered the impedance further, as noted by Marmont [2]. A second layer of silver chloride did not result in any further reduction in impedance, as hypothesized.
Keywords
Biomedical electrodes; Biomedical engineering; Biomedical measurements; Circuits; Coatings; Impedance measurement; Silver; Skin; Surface impedance; Surface treatment; Chlorides; Electricity; Electrodes; Silver; Surface Properties;