    An In Vivo Study of Cardiac Pacemaker Optimization by Pulse Shape Modification

  • Author

    Klafter, Richard D. ; Hrebien, Leonid

  • Author_Institution
    Biomedical Engineering and Science Program, Drexel University, Philadelphia, PA 19104.
  • Issue
    3
  • fYear
    1976
  • fDate
    5/1/1976 12:00:00 AM
  • Firstpage
    233
  • Lastpage
    239
  • Abstract
    This paper is an experimental extension of theoretical work previously published by one of the authors. Results of pacing a canine heart using four different increasing exponential and ramp-like waveforms are presented. It is shown that these waveforms reduce the cardiac threshold energy by as much as 14 percent when compared with a pure rectangular pulse shape and by about 21 percent when compared with a standard cardiac pacer pulse. It is also found that above threshold energy is reduced by 50-60 percent. These results offer promise of increasing the life of certain new pacer power sources currently being tested. In addition, it may now be possible to pace at lower energy levels, thereby reducing tissue damage near the stimulating electrode, without running as much of a risk of exit block as with a comparable rectangular pulse pacer.
  • Keywords
    Energy states; Heart; In vivo; Life testing; Pacemakers; Pulse shaping methods; Pulsed power supplies; Shape; Space vector pulse width modulation; Tissue damage; Animals; Dogs; Electronics, Medical; Models, Biological; Pacemaker, Artificial;