DocumentCode
1208132
Title
Peak Efficiency of NaI(Tl) Crystals for Gamma Rays from 0.150 to 7.5 Mev
Author
Lazar, N.H. ; Davis, R.C. ; Bell, P.R.
Author_Institution
Oak Ridge Natl. Lab., Oak Ridge, Tenn.
Volume
3
Issue
4
fYear
1956
Firstpage
136
Lastpage
137
Abstract
The peak efficiencies of cylindrical 1-1/2 X 1-inch, 3 X 3-inch and 3 X 3-inch bevelled crystals have been determined by the multiplication of the theoretical total efficiency for a particular energy gamma ray by the experimentally determined ratio of the areas under the full energy peak and under the total pulse height spectrum for the gamma ray. Care was taken to eliminate spurious effects in the pulse height spectrum such as scattering, bremsstrahlung, and other interfering gamma radiation from the source, etc., in the experimental measurements. The peak efficiencies have also been determined at certain energies by Ã-¿ coincidence experiments and yield the same results within 5 per cent for energies to 2.76 mev. The intensities of gamma rays determined in this way have been compared to the disintegration rates obtained from 4¿ Ã-counting techniques from sources with known decay schemes, and the results agreed within the limits given above. The extension of these efficiencies to crystals of different composition or to crystals with the same shapes but different dimensions will be discussed.
Keywords
Crystals; Face detection; Gamma rays; Geometry; Nuclear and plasma sciences; Pulse measurements; Radiation detectors; Scattering; Shape measurement; Solids;
fLanguage
English
Journal_Title
Nuclear Science, IRE Transactions on
Publisher
ieee
ISSN
0096-2015
Type
jour
DOI
10.1109/TNS2.1956.4315564
Filename
4315564
Link To Document