    Evaluation of the Parallel Conductor Theory for Measuring Human Limb Blood Flow by Electrical Admittance Plethysmography

  • Author

    Shimazu, Hideaki ; Yamakoshi, Ken-ichi ; Togawa, Tatsuo ; Fukuoka, Masakazu ; Ito, Hiroshi

  • Author_Institution
    Research Institute of Applied Electricity, Hokkaido University
  • Issue
    1
  • fYear
    1982
  • Firstpage
    1
  • Lastpage
    7
  • Abstract
    We devised a compensation technique that measures the admittance change of a limb submerged in an electrolyte solution in a cylinder. Using this technique, we evaluated the accuracy of admittance plethysmography and the validity of the parallel-conductor model on which the theory of blood flow measurement by electrical admittance (or impedance) plethysmography is based. From a theoretical point of view, if a limb is regarded as a parallel-conductor model, the admittance change due to blood pooling following venous occlusion should disappear when the resistivity of the solution is equal to that of the blood.
  • Keywords
    Admittance measurement; Anthropometry; Blood flow; Conductivity; Conductors; Electric variables measurement; Fluid flow measurement; Humans; Impedance measurement; Plethysmography; Blood Flow Velocity; Evaluation Studies as Topic; Extremities; Humans; Plethysmography, Impedance; Regional Blood Flow;