• DocumentCode
    1209798
  • Title

    An Amplitude and Phase Regulating Magnetic Field Generator for an Eye Movement Monitor

  • Author

    Optican, L.M. ; Frank, D.E. ; Smith, B.M. ; Colburn, Andt R.

  • Author_Institution
    Laboratory of Sensorimotor Research, National Eye Institute, National Institutes of Health
  • Issue
    3
  • fYear
    1982
  • fDate
    3/1/1982 12:00:00 AM
  • Firstpage
    206
  • Lastpage
    209
  • Abstract
    One common method of eye movement detection uses an ocular search coil within two magnetic fields arranged in spatial and temporal quadrature. An inexpensive circuit is presented that regulates the strength of the two fields and their phase separation. This stabilization minimizes perturbations due to motion of the field´s generating coils, motion of metal witiin or near the fi´elds, and component value drift.
  • Keywords
    Bandwidth; Circuits; Coils; Decoding; Electromagnetic fields; Laboratories; Magnetic fields; Monitoring; Phase detection; Voltage; Calibration; Electromagnetic Fields; Electromagnetics; Electronystagmography; Equipment Design; Humans;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.1982.324888
  • Filename
    4121386