            Abstract : 
A method for making high-speed measurements of sarcomere length directly from the muscle striation image is presented. A phase-locked loop is used with a self-scanned photodiode array to measure the spatial frequency of the striation pattern. Computation circuitry converts frequency to sarcomere length. The sarcomere length output has a resolution better than 2 nm at a sample rate of once every 250¿s.
         
        
            Keywords : 
Circuits; Frequency measurement; Length measurement; Muscles; Phase locked loops; Phase measurement; Phased arrays; Photodiodes; Spatial resolution; Time measurement; Animals; Electronics, Medical; Microscopy; Myofibrils; Regression Analysis;
         
        
        