DocumentCode
1210611
Title
Solutions to Three Capacitance Problems in Intracellular Single-Electrode Bridge Amplifiers
Author
Fox, Steven E.
Author_Institution
Department of Physiology, Downstate Medical Center
Issue
9
fYear
1982
Firstpage
662
Lastpage
666
Abstract
Measurements of membrane impedance using single-electrode intracellular bridge amplifiers may be seriously distorted even at frequencies as low as 16 Hz by capacitances associated with the current limiting resistor and the electrode. Modifications of the capacity compensation circuit are outlined, which eliminate much of this error, allowing measurements with 108 a conventional micropipets at frequencies as high as 400 Hz.
Keywords
Biomembranes; Bridge circuits; Capacitance measurement; Current limiters; Current measurement; Distortion measurement; Electrodes; Frequency measurement; Impedance measurement; Resistors; Electric Conductivity; Electronics, Medical; Membrane Potentials;
fLanguage
English
Journal_Title
Biomedical Engineering, IEEE Transactions on
Publisher
ieee
ISSN
0018-9294
Type
jour
DOI
10.1109/TBME.1982.324939
Filename
4121488