DocumentCode
1210671
Title
Recognition and Storage of Metal Heart Marker Position from Biplane X-Ray Images at Video Rates
Author
Philips, Charles M. ; Prenis, John ; Santamore, William P. ; Bove, Alfred A.
Author_Institution
Department of Medical Physics, School of Medicine, Temple University
Issue
1
fYear
1983
Firstpage
10
Lastpage
17
Abstract
We have developed a semiautomatic microcomputer-based instrument for recording the three-dimensional positions of metal heart markers from biplane X-ray images at 60 frames/s. This instrument eliminates the need for film processing and tedious digitizing by human operators. Markers are recognized electronically on each video line using criteria of density changes (edge detection), size, and location. Only the X-Y or Z-Y location of each marker is stored. This transformation of the gray scale image into a list of coordinates achieves a data compression of 250 fold. A small fast memory and a second large one provide an effective slowing of the high peak rates of the marker location data. This buffering allows the marker coordinates to be permanently stored on a floppy disk for later analysis. Four s of data can be stored in one burst when 20 markers are used. The instrument was designed to measure the pumping action of the heart through the calculation of instantaneous local strain, wall thickness, and chamber volume. This paper compares alternate recognition strategies and storage methods.
Keywords
Buffer storage; Heart; Humans; Image recognition; Image storage; Instruments; Strain measurement; Video compression; Video recording; X-ray imaging; Animals; Dogs; Electronics, Medical; Heart; Microcomputers;
fLanguage
English
Journal_Title
Biomedical Engineering, IEEE Transactions on
Publisher
ieee
ISSN
0018-9294
Type
jour
DOI
10.1109/TBME.1983.325161
Filename
4121497
Link To Document