• DocumentCode
    1210735
  • Title

    Reductionl of Interference Due to Common Mode Voltage in Biopotential Amplifiers

  • Author

    Winter, Bruce B. ; Webster, John G.

  • Author_Institution
    IBM Corporation
  • Issue
    1
  • fYear
    1983
  • Firstpage
    58
  • Lastpage
    62
  • Abstract
    We review how the nonideal properties of biopotential amplifiers can transform common mode voltage into interference. We then review several design approaches for two-and three-electrode amplifiers, both nonisolated and isolated, that reduce this interference. We consider the effects of static electricity on the various designs, and we show how to calculate the optimal values of the circuit components.
  • Keywords
    Capacitance; Circuits; Differential amplifiers; Electrocardiography; Electrodes; Electroencephalography; Impedance; Interference; Magnetic recording; Voltage; Electrocardiography; Electroencephalography; Electromagnetic Fields; Electromyography; Electronics, Medical; Humans;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.1983.325167
  • Filename