DocumentCode
1210870
Title
Systematic and Random Errors in the Determination of Respiratory Impedance by Means of the Forced Oscillation Technique: A Theoretical Study
Author
Franken, H. ; Clément, J. ; Van de Woestijne, K.P.
Author_Institution
Laboratorium voor Longfunktieonderzoek, Academisch Ziekenhuis St. Rafael
Issue
10
fYear
1983
Firstpage
642
Lastpage
651
Abstract
The measurement of the impedance of the respiratory system by means of the forced oscillation technique is disturbed by the breathing signal of the subject. The latter introduces systematic and random errors on the obtained impedance values. The size of these errors depends on the relative power of the breathing signal with respect to the forced oscillatory signal x2.
Keywords
Fluid flow measurement; Force measurement; Frequency; Impedance measurement; Loudspeakers; Measurement standards; Noise measurement; Pressure measurement; Respiratory system; Zirconium; Airway Resistance; Humans; Lung Diseases, Obstructive; Methods; Respiratory Physiology;
fLanguage
English
Journal_Title
Biomedical Engineering, IEEE Transactions on
Publisher
ieee
ISSN
0018-9294
Type
jour
DOI
10.1109/TBME.1983.325066
Filename