DocumentCode
1211104
Title
An effective compensation method for the mutual coupling effect in phased arrays for magnetic resonance imaging
Author
Hui, Hon Tat
Author_Institution
Sch. of Inf. Technol. & Electr. Eng., Univ. of Queensland, Brisbane, Australia
Volume
53
Issue
11
fYear
2005
Firstpage
3576
Lastpage
3583
Abstract
An effective compensation method to compensate for the mutual coupling effect in magnetic resonance imaging (MRI) phased arrays is introduced. This method uses the knowledge of the position of the signal source in MRI, i.e., the active slice, to define a new mutual impedance that accurately quantifies the coupled voltages and enables them to be removed from the terminal voltages almost completely. Numerical results using the method of moments show that the percentage errors in the compensated voltage are at least on the order of 10-5% and the isolations between two coils are more than 120 dB even at a low-field case of 0.5 T (f0=21.3 MHz). This method can be implemented by either software or hardware.
Keywords
antenna phased arrays; biomedical MRI; electromagnetic coupling; method of moments; signal sources; MRI; compensation method; magnetic resonance imaging; method of moment; mutual coupling effect; mutual impedance; phased array; signal source; voltage coupling; Antenna arrays; Coils; Current distribution; Impedance; Magnetic resonance imaging; Moment methods; Mutual coupling; Phased arrays; Preamplifiers; Voltage; Compensation method; magnetic resonance imaging (MRI); mutual coupling; mutual impedance; phased array;
fLanguage
English
Journal_Title
Antennas and Propagation, IEEE Transactions on
Publisher
ieee
ISSN
0018-926X
Type
jour
DOI
10.1109/TAP.2005.858867
Filename
1528726
Link To Document