• DocumentCode
    1211104
  • Title

    An effective compensation method for the mutual coupling effect in phased arrays for magnetic resonance imaging

  • Author

    Hui, Hon Tat

  • Author_Institution
    Sch. of Inf. Technol. & Electr. Eng., Univ. of Queensland, Brisbane, Australia
  • Volume
    53
  • Issue
    11
  • fYear
    2005
  • Firstpage
    3576
  • Lastpage
    3583
  • Abstract
    An effective compensation method to compensate for the mutual coupling effect in magnetic resonance imaging (MRI) phased arrays is introduced. This method uses the knowledge of the position of the signal source in MRI, i.e., the active slice, to define a new mutual impedance that accurately quantifies the coupled voltages and enables them to be removed from the terminal voltages almost completely. Numerical results using the method of moments show that the percentage errors in the compensated voltage are at least on the order of 10-5% and the isolations between two coils are more than 120 dB even at a low-field case of 0.5 T (f0=21.3 MHz). This method can be implemented by either software or hardware.
  • Keywords
    antenna phased arrays; biomedical MRI; electromagnetic coupling; method of moments; signal sources; MRI; compensation method; magnetic resonance imaging; method of moment; mutual coupling effect; mutual impedance; phased array; signal source; voltage coupling; Antenna arrays; Coils; Current distribution; Impedance; Magnetic resonance imaging; Moment methods; Mutual coupling; Phased arrays; Preamplifiers; Voltage; Compensation method; magnetic resonance imaging (MRI); mutual coupling; mutual impedance; phased array;
  • fLanguage
    English
  • Journal_Title
    Antennas and Propagation, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-926X
  • Type

    jour

  • DOI
    10.1109/TAP.2005.858867
  • Filename
    1528726