DocumentCode :
1211866
Title :
Effects of Diffraction on Measurements of the Frequency-Dependent Ultrasonic Attenuation
Author :
Donnell, Matthew O.
Author_Institution :
Corporate Research and Development, General Electric Company
Issue :
6
fYear :
1983
fDate :
6/1/1983 12:00:00 AM
Firstpage :
320
Lastpage :
326
Abstract :
Using a one-dimensional model of wave propagation, the ultrasonic attenuation can be estimimated from spectral analysis of RF A-scans. The results of computer modeling studies and measurements on tissue equivalent phantoms indicate that spectral analysis methods are accurate only in the regime where a one-dimensional model is adequate. In the near field of a typical B-scan transducer, a one-dimenional model is not an accurate description of wave propagation. A simple processing method is presented that minimizes near field diffraction effects on estimates of the slope of the attenuation coefficient as a function of frequency.
Keywords :
Apertures; Attenuation measurement; Diffraction; Frequency estimation; Frequency measurement; Radio frequency; Signal processing; Spectral analysis; Ultrasonic transducers; Ultrasonic variables measurement; Computers; Models, Biological; Models, Structural; Spectrum Analysis; Transducers; Ultrasonics;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/TBME.1983.325131
Filename :
4121641
Link To Document :
بازگشت