DocumentCode
1212276
Title
LRR-a self-calibration technique for the calibration of vector network analyzers
Author
Rolfes, Ilona ; Schiek, Burkhard
Author_Institution
RF & Microwave Eng. Inst., Ruhr-Univ. of Bochum, Germany
Volume
52
Issue
2
fYear
2003
fDate
4/1/2003 12:00:00 AM
Firstpage
316
Lastpage
319
Abstract
A new self-calibration procedure, the LRR method, for the calibration of vector network analyzers is presented. The calibration circuits consist of partly unknown standards, where L symbolizes a line element and R represents a symmetrical reflection standard. In contrast to the thru reflect line (TRL) method, which needs a line-standard with a different length than the other calibration standards, the calibration circuits of the LRR method are all of equal mechanical length. This is advantageous because the connectors of the vector network analyzer do not, thus, have to be placed at different distances from each other during calibration. In addition, the complexity of the test fixture can be reduced. The robust functionality of the LRR method is confirmed by measurements.
Keywords
S-parameters; calibration; measurement standards; microwave measurement; network analysers; LRR method; VNA; calibration circuits; measurement standards; microwave measurements; scattering parameters measurement; self-calibration procedure; symmetrical reflection standard; test fixture complexity; vector network analyzers; Calibration; Circuit testing; Connectors; Fixtures; Microwave measurements; Reflection; Robustness; Space technology; Symmetric matrices; Transmission line matrix methods;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/TIM.2003.810040
Filename
1202037
Link To Document