Title :
Calibration of electronic capacitance and dissipation factor bridges
Author :
Ramm, Günther ; Moser, Harald
Author_Institution :
Phys. Tech. Bundesanstalt, Braunschweig, Germany
fDate :
4/1/2003 12:00:00 AM
Abstract :
A new calibration procedure for electronic capacitance and dissipation factor bridges is presented. These bridges measure the ratio of their input currents. For the calibration, two sinusoidal alternating currents are synthesized with the aid of a digital, programmable two-channel AC voltage source and two AC resistors. The complex ratio of these synthesized input currents is calculated from an AC voltage ratio determined by a sampling method, together with the impedance ratio of the AC resistors. At currents of a few mA, their complex ratio in the frequency range up to 100 Hz can be measured with uncertainties (k=2) of approximately 1×10-6 and 1 μrad, respectively.
Keywords :
bridge circuits; bridge instruments; calibration; capacitance measurement; electric impedance; measurement uncertainty; programmable circuits; signal sampling; voltage measurement; 100 Hz; AC bridge; AC resistor impedance ratio; calibration; complex voltage ratio; digital programmable two-channel AC voltage source; electronic capacitance/dissipation factor bridges; frequency range; impedance ratio; input current ratio measurement; measurement uncertainties; sampling voltmeter; sinusoidal alternating current synthesis; Bridge circuits; Calibration; Capacitance; Current measurement; Frequency measurement; Impedance; Measurement uncertainty; Resistors; Sampling methods; Voltage;
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
DOI :
10.1109/TIM.2003.810719