DocumentCode :
1212583
Title :
Multiple Sensor Optical Thermometry System for Application in Clinical Hyperthermia
Author :
Vaguine, Victor A. ; Christensen, Douglas A. ; Lindley, Joe H. ; Walston, Thomas E.
Author_Institution :
Clini-Therm Corporation
Issue :
1
fYear :
1984
Firstpage :
168
Lastpage :
172
Abstract :
The thermometry system described is based upon the temperature dependence of the band edge absorption of infrared light in GaAs crystal. The design of the thermometry was completed, and the system was subjected to an extensive evaluation, including testing with tissue phantoms and microwave applicators. The system has up to 12 temperature sensors which are packaged in two basic probe coinfigurations: a single-sensor probe with a length of 1.2 m and a diameter of 0.6 mm; and a four-sensor linear array probe with a length of 1.2 m, diameter of 1.1 mm, and spacing of 1.5 cm between adjacent sensors. Results of thermometry evaluation are presented, including data on automatic calibration, temperature accuracy and stability, and EMI protection.
Keywords :
Electromagnetic wave absorption; Gallium arsenide; Hyperthermia; Optical sensors; Probes; Sensor arrays; Sensor systems and applications; Temperature dependence; Temperature sensors; Thermal sensors; Equipment Design; Fiber Optics; Humans; Hyperthermia, Induced; Microcomputers; Thermometers;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/TBME.1984.325383
Filename :
4121727
Link To Document :
بازگشت