A precision setup and method for calibrating a DC voltage divider´s ratios from 10 V to 1000 V
Author :
Slinde, Harald ; Lind, Kåre
Author_Institution :
Lab. Dept., Justervesenet, Kjeller, Norway
Volume :
52
Issue :
2
fYear :
2003
fDate :
4/1/2003 12:00:00 AM
Firstpage :
461
Lastpage :
464
Abstract :
Resistive dividers are often used to calibrate DC voltage calibrators and voltmeters above 10 V, with ratio uncertainties lower than 10-6. The dividers need therefore to be calibrated with uncertainties lower than 5×10-7. A setup and method for the calibration of a resistive divider at its working voltage has been developed. The repeatability was found to be good enough to give standard deviations of the mean lower than ±1×10-8 in the measured ratio 1000/10. The total uncertainties were estimated to be less than ±1.5×10-7 (k=1) for all measured ratios.
Keywords :
calibration; measurement uncertainty; voltage dividers; voltage measurement; 10 to 1000 V; DC voltage divider calibration; international comparison; precision setup; repeatability; resistive dividers; traceability; voltage measurements; Calibration; Detectors; Helium; Instruments; Low pass filters; Measurement standards; Resistors; Uncertainty; Voltage measurement; Voltmeters;