DocumentCode
1212804
Title
Evaluation of Light-Emitting Diodes for Whole Blood Oximetry
Author
Shepherd, A.P. ; Kiel, J.W. ; Riedel, G.L.
Author_Institution
Department of Physiology, University of Texas Health Science Center
Issue
11
fYear
1984
Firstpage
723
Lastpage
725
Abstract
The difference between the optical extinction coefficients of oxy-and deoxyhemoglobin is a primary concern in the design of blood oxygen instrumentation. In this communication, we report the apparent values for the molar extinction coefficients of hemoglobin under conditions of interest to instrument designers, namely, with whole blood and with unfiltered LED illumination. Extinction coefficient values are reported for four commercially available LED´s with typical spectral half-intensity bandwidths and peak emissions ranging from 660 to 940 nm.
Keywords
Biomedical computing; Blood; Electroencephalography; Frequency conversion; Gaussian distribution; Light emitting diodes; Optical noise; Parametric statistics; Signal analysis; Time domain analysis; Biomedical Engineering; Humans; Light; Oximetry; Oxygen;
fLanguage
English
Journal_Title
Biomedical Engineering, IEEE Transactions on
Publisher
ieee
ISSN
0018-9294
Type
jour
DOI
10.1109/TBME.1984.325398
Filename
4121757
Link To Document