DocumentCode :
1212874
Title :
Direct determination of capacitance standards using a quadrature bridge and a pair of quantized Hall resistors
Author :
Inglis, A.D. ; Wood, B.M. ; Côté, M. ; Young, R.B. ; Early, M.D.
Author_Institution :
Inst. for Nat. Meas. Stand., Nat. Res. Council, Ottawa, Ont., Canada
Volume :
52
Issue :
2
fYear :
2003
fDate :
4/1/2003 12:00:00 AM
Firstpage :
559
Lastpage :
562
Abstract :
We have used a quadrature bridge to compare directly two quantum Hall resistors with two 10-nF capacitors at 1233 Hz. This approach eliminates uncertainties due to thermal and structural instabilities and to AC/DC differences of the wire resistors normally used in an intermediate step. We have measured the capacitative interaction between the devices, and have developed a shielded header to suppress this effect. The value for the National Research Council primary standard 10-pF capacitor obtained by this method agrees with the expected value to within 0.1±3.5 aF, or (0.01±0.35)×10-6.
Keywords :
Hall effect devices; bridge circuits; capacitance measurement; electromagnetic shielding; measurement standards; quantum Hall effect; resistors; thermal stability; 10 nF; 1233 Hz; AC bridges; National Research Council primary standard capacitor; capacitative interaction; capacitors; direct capacitance standards determination; quadrature bridge; quantized Hall resistor pair; shielded header; structural instabilities; thermal instabilities; Bridge circuits; Capacitance measurement; Capacitors; Councils; Laboratories; Measurement standards; Quantum capacitance; Resistors; Voltage; Wire;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.2003.810454