• DocumentCode
    1213517
  • Title

    An Inexpensive Eye Movement Monitor Using the Scleral Search Coil Technique

  • Author

    Remmel, Ronald S.

  • Author_Institution
    Department of Biomedical Engineering, Boston University
  • Issue
    4
  • fYear
    1984
  • fDate
    4/1/1984 12:00:00 AM
  • Firstpage
    388
  • Lastpage
    390
  • Abstract
    For the economical recording of eye movements using the scleral search-coil method, we developed a circuit using common components and costing only $300. The Helmholtz field coils are 51 cm in diameter and are separated by 51 cm. The horizontal and vertical field coils are driven at 50 and 75 kHz, respectively, by phase-locked square waves; no tuning is required. After amplification, the eye-coil signal is phase detected at 50 and at 75 kHz to produce the horizontal and vertical eye-position signals, respectively. For a 2 cm diameter eye coil of one turn, the noise is 1.0 min of angle rms.
  • Keywords
    Circuits; Coils; Frequency; Magnetic flux; Monitoring; Phase detection; Shift registers; Switches; Voltage; Wire; Biomedical Engineering; Eye Movements; Models, Theoretical;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.1984.325352
  • Filename