DocumentCode
1214250
Title
Analysis of a Digital EMG Signal Processor in Dynamic Conditions
Author
Alessio, T.D.
Author_Institution
Dipartimento di Scienza e Tecnica dell´´lnformazione e della Comunicazione, Universiti di Roma
Issue
1
fYear
1985
Firstpage
78
Lastpage
82
Abstract
The performances of digital electromyographic signal processors in dynamic conditions are determined by evaluating the root-mean-square value of the estimation error, including bias and variance of the estimates. The results thus obtained are valid for a wide class of myoprocessor schemes. The analysis has been used to optimize myoprocessors with moving-average smoothing filters, the length of which will be adapted to the local variability characteristics of the force in order to obtain a minimum overali estimation error.
Keywords
Data acquisition; Electromyography; Fatigue; Frequency; Humans; Magnetic force microscopy; Monitoring; Signal analysis; Signal processing; White noise; Biomedical Engineering; Electromyography; Humans;
fLanguage
English
Journal_Title
Biomedical Engineering, IEEE Transactions on
Publisher
ieee
ISSN
0018-9294
Type
jour
DOI
10.1109/TBME.1985.325635
Filename
4121932
Link To Document