An Improved Method for Detection of Hemagglutination Using an Automated Microplate Reader
Author
Severns, Matthew L. ; Kline, Linda M.
Author_Institution
Biomedical Engineering Laboratory, American Red Cross Blood Services Laboratories
Issue
5
fYear
1985
fDate
5/1/1985 12:00:00 AM
Firstpage
349
Lastpage
352
Abstract
Several new automated microplate readers which can take multiple readings in each well have, been commercially developed. Data gathered from one of these readers have been analyzed to determine a parsimonious algorithm for discriminating positive hemagglutination reactions from negative ones. The resulting classification algorithm provides more accurate classification of weak positive reactions than the one point measurements normally used with only a moderate increase in computational complexity.
Keywords
Algorithm design and analysis; Automatic testing; Biomedical measurements; Cells (biology); Performance evaluation; Plasmas; Position measurement; Red blood cells; System testing; Wavelength measurement; Autoanalysis; Blood Grouping and Crossmatching; Hemagglutination Tests; Humans;