• DocumentCode
    1215741
  • Title

    Fast Response Ultrasonic Flowmeter Measures Breathing Dynamics

  • Author

    Lynnworth, Lawrence C. ; Korba, James M. ; Wallace, D.Russell

  • Author_Institution
    Panametrics, Inc.
  • Issue
    7
  • fYear
    1985
  • fDate
    7/1/1985 12:00:00 AM
  • Firstpage
    530
  • Lastpage
    535
  • Abstract
    Questions about the dynamics of respiration following prolonged exposure to hypogravity led to a set of flowmeter requirements that included: ~240 Hz sampling rate, ±0.02 1/s accuracy, 500: 1 bidirectional range (±0.02 to ±10 1/s), sample volume ¿50 cm3, and small pressure drop. In the present laboratory feasibility study, the investigators modified an industrial ultrasonic flowmeter such that 100 kHz pulses, alternately transmitted upstream and downstream at a 300 Hz rate, interrogated inhaled or exhaled air flowing through a 2 ×2 cm square duct or a 1.27 ×2.54 cm rectangular duct. This communication emphasizes the electronic and programming aspects of the ultrasonic flowmeter developed for this study, and includes waveform illustrations and upstream-downstream time difference data obtained in laboratory tests.
  • Keywords
    Atmospheric measurements; Biomedical measurements; Ducts; Fluid flow; Fluid flow measurement; Laboratories; Sampling methods; Time measurement; Ultrasonic variables measurement; Volume measurement; Biomedical Engineering; Humans; Respiration; Rheology; Ultrasonics;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.1985.325570
  • Filename
    4122104