Title :
On the Reduction of Interference Due to Common Mode Voltage in Two-Electrode Biopotential Amplifiers
Author :
Pallas-areny, Ramon
Author_Institution :
Laboratorio de Instrumentación y Bioingenieria, Escuela Técnica Superior Ingenieros Telecomunicaci6n
Abstract :
The circuit parameters contributing to common mode voltage interferences in two-electrode amplifiers, both isolated and nonisolated, are reviewed. For the isolated case, a more complete model than that reported in the literature is described, and a new methodology to analyze the problem is introduced. Next, the usual values of the parameters concerned are considered; and by applying the results obtained with the model and method described, some conclusions about the design of this kind of amplifier are drawn.
Keywords :
Capacitance; Circuits; Disk recording; Electrocardiography; Electrodes; Impedance; Interference; Operational amplifiers; Telemetry; Voltage; Amplifiers; Electrocardiography; Electrodes; Electroencephalography; Humans;
Journal_Title :
Biomedical Engineering, IEEE Transactions on
DOI :