  • Abstract
    In biomedical signal averaging applications where a small repetitive signal is to be extracted form a very noisy waveform (noise variance ¿2n), the A/D converter range is set at ±A¿n where A typically has a value of 3 or 4. In this case, A/D roundoff noise using a (b + 1)-bit A/D converter degrades the SNR of the resulting signal estimate by an amount slightly greater than 10 log (1 + A2¿2b/12). Application of this result to the design of a system for digital acquisition and averaging of the acoustic brainstem response indicates that an increase of A/D resolution beyond the first few bits yields little improvement in the SNR of the measurement.
  • Keywords
    Acoustic noise; Additive noise; Degradation; Noise reduction; Quantization; Random processes; Roundoff errors; Signal processing; Signal sampling; Signal to noise ratio; Biomedical Engineering; Brain Stem; Humans;