Title :
Side-opened out-of-plane microneedles for microfluidic transdermal liquid transfer
Author :
Griss, Patrick ; Stemme, Göran
Author_Institution :
Dept. of Signals, Sensors & Syst., R. Inst. of Technol., Stockholm, Sweden
fDate :
6/1/2003 12:00:00 AM
Abstract :
We present the first hollow out-of-wafer-plane silicon microneedles having openings in the shaft rather than having an orifice at the tip. These structures are well suited for transdermal microfluidic applications, e.g., drug or vaccine delivery. The developed deep-reactive ion etching (DRIE) process allows fabrication of two dimensional, mechanically highly resistant needle arrays offering low resistance to liquid flows and a large exposure area between the fluid and the tissue. The presented process does not require much wafer handling and only two photolithography steps are required. Using a 3×3 mm2 chip in a typical application, e.g., vaccine delivery, a 100 μl volume of aqueous fluid injected in 2 s would cause a pressure drop of less than 2 kPa. The presented needles are approximately 210 μm long.
Keywords :
biomedical electronics; drug delivery systems; microfluidics; photolithography; sputter etching; 210 micron; Si; aqueous fluid; deep-reactive ion etching; exposure area; hollow out-of-wafer-plane microneedles; liquid flows; microfluidic transdermal liquid transfer; photolithography; pressure drop; side-opened out-of-plane microneedles; vaccine delivery; Drugs; Etching; Fabrication; Fluid flow; Microfluidics; Needles; Orifices; Shafts; Silicon; Vaccines;
Journal_Title :
Microelectromechanical Systems, Journal of
DOI :
10.1109/JMEMS.2003.809959