• DocumentCode
    1218327
  • Title

    A Microprocessor-Based Two-Channel Thromboelastograph

  • Author

    Chen, Shoupu ; Thakor, Nitish V. ; Wagner, James W.

  • Author_Institution
    Department of Biomedical Engineering, Johns Hopkins School of Medicine
  • Issue
    9
  • fYear
    1986
  • Firstpage
    887
  • Lastpage
    890
  • Abstract
    This paper presents a design for a digital thromboelastograph (TEG). The instrument design is based on the principle of a couette viscometer. The device measures shear elasticity of blood during clot formation. The recording mechanism comprises a two-channel light-beam deflection system and a recording system based on a charge-coupled-device (CCD) array. An 8-bit CMOS microprocessor (NSC800) supervises data acquisition from the CCD array. The two-channel system produces a record on a thermal printer allowing one to observe the time course of coagulation and simultaneously compare two different blood specimens.
  • Keywords
    Biomedical measurements; Blood; Charge coupled devices; Coagulation; Engine cylinders; Mirrors; Optical arrays; Optical recording; Viscosity; Wire; Animals; Blood Coagulation; Computers; Dogs; Heparin; Microcomputers; Thrombelastography;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.1986.325784
  • Filename
    4122415