• DocumentCode
    1218603
  • Title

    Ultrasound Doppler Probing of Flows Transverse with Respect to Beam Axis

  • Author

    Newhouse, Vernon L. ; Censor, Dan ; Vontz, Thomas ; Cisneros, Jose A. ; Goldberg, Barry B.

  • Author_Institution
    Biomedical Engineering and Science Institute and the Department of Electrical and Computer Engineering, Drexel University
  • Issue
    10
  • fYear
    1987
  • Firstpage
    779
  • Lastpage
    789
  • Abstract
    It is an accepted fact that transverse Doppler effects of the first order in v/c are nonexistent for all physical wave phenomena, including acoustics, i.e., the Doppler effect is zero for radiation normal to the direction of motion. However, this statement assumes that the incident field is a plane wave, which is not true in general for finite aperture sources. Consequently, the probing of flows transverse to the axis of finite diameter beams, particularly focused beams, is feasible. This geometry will be advantageous in many applications where the classical orientation of the sound beam, oblique to the flow, is not possible. With this motivation in mind, the theory and experimental feasibility of measuring Doppler spectra in transverse geometries is presently investigated.
  • Keywords
    Acoustic beams; Acoustic waves; Biomedical engineering; Biomedical measurements; Doppler effect; Electromagnetic scattering; Fluid flow measurement; Pulse measurements; Ultrasonic imaging; Ultrasonic variables measurement; Pulsatile Flow; Rheology; Ultrasonics;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.1987.325920
  • Filename
    4122448