DocumentCode :
1219035
Title :
Development of an Active Intravascular MR Device With an Optical Transmission System
Author :
Fandrey, Stephan ; Weiss, Steffen ; Müller, Jörg
Author_Institution :
Inst. of Micro Syst. Technol., Hamburg Univ. of Technol., Hamburg
Volume :
27
Issue :
12
fYear :
2008
Firstpage :
1723
Lastpage :
1727
Abstract :
Magnetic resonance imaging (MRI) is a safe and reliable medical imaging method providing good soft tissue contrast while avoiding harmful ionizing radiation. It is highly desirable to use the MRI technology for interventional procedures. However, due to resonance effects that can result in tissue heating, long conducting cables must be avoided. Motivated by the need for more radio-frequency (RF) safety, we developed an optical transmission system for active intravascular MRI devices. An optical transmitter sends the MR signal via an optical fiber. A miniature optical modulator was designed to be integrated into a catheter tip. Furthermore, power is supplied optically to the transmitter. This system can target new medical applications, due to safe catheter tracking and safe intravascular imaging.
Keywords :
biomedical MRI; biomedical equipment; blood vessels; catheters; optical fibres; optical modulation; optical transmitters; patient care; active intravascular MR device; catheter; magnetic resonance imaging; miniature optical modulator; optical fiber; optical power supply; optical transmission system; optical transmitter; radio-frequency safety; soft tissue contrast; tissue heating; Biological tissues; Biomedical imaging; Biomedical optical imaging; Catheters; Ionizing radiation; Magnetic resonance imaging; Optical devices; Optical modulation; Optical transmitters; Radio frequency; Catheter; catheter; intravascular imaging; magnetic resonance imaging (MRI); optical power supply; optical transmission; radio-frequency (RF) safety; rf-safety; Catheterization; Equipment Design; Fiber Optic Technology; Humans; Magnetic Resonance Angiography; Optical Fibers; Phantoms, Imaging; Safety;
fLanguage :
English
Journal_Title :
Medical Imaging, IEEE Transactions on
Publisher :
ieee
ISSN :
0278-0062
Type :
jour
DOI :
10.1109/TMI.2008.925079
Filename :
4520151
Link To Document :
بازگشت