• DocumentCode
    1219240
  • Title

    A Comparison of Finite Element and Integral Equation Formulations for the Calculation of Electrocardiographic Potentials-II

  • Author

    Pilkington, Theo C. ; Morrow, Mary N. ; Stanley, Peggy C.

  • Author_Institution
    Department of Biomedical Engineering, Duke University
  • Issue
    3
  • fYear
    1987
  • fDate
    3/1/1987 12:00:00 AM
  • Firstpage
    258
  • Lastpage
    260
  • Abstract
    This communication is an extension of previous work which reported that the use of a basic finite element formulation improves the accuracy over that obtainable with integral equations, and that the improvement in accuracy is particularly notable for inverse estimation. Recent results indicate that a judicious choice of triangular subdivisions combined with the utilization of Radon integration to evaluate the contributions of each of these smaller triangles produces integral equation formulations with accuracy roughly comparable to that obtained using finite element formulations.
  • Keywords
    Biomedical computing; Biomedical engineering; Conductivity; Finite element methods; Geometry; Integral equations; Low pass filters; Modems; Solids; Voltage; Electrocardiography; Humans; Mathematics;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.1987.325956
  • Filename
    4122521