• DocumentCode
    1219442
  • Title

    An advanced signal processing technique for impedance cardiography

  • Author

    Xiang Wang ; Sun, H.H. ; Van de Water, J.M.

  • Author_Institution
    Renaissance Technol. Inc., Newton, PA, USA
  • Volume
    42
  • Issue
    2
  • fYear
    1995
  • Firstpage
    224
  • Lastpage
    230
  • Abstract
    A new design using the latest technique in signal processing, the time-frequency analysis method, was developed to process impedance cardiography signals. This technique, when used to determine the relevant calculation parameters, was found to be more accurate than conventional methods. It was shown to be advantageous in reducing ventilation artifacts and motion noise, resulting in greater accuracy. Its cardiac output values had a much better correlation coefficient when compared in the clinical setting to the standard thermodilution technique than did the values from conventional impedance cardiography devices.
  • Keywords
    cardiology; electric impedance imaging; medical signal processing; time-frequency analysis; advanced signal processing technique; calculation parameters; cardiac output values; correlation coefficient; impedance cardiography signals; motion noise; time-frequency analysis method; ventilation artifacts; Biomedical signal processing; Cardiography; Electrodes; Impedance; Signal design; Signal processing; Sun; Thorax; Time frequency analysis; Ventilation; Cardiography, Impedance; Electrodes; Equipment Design; Humans; Signal Processing, Computer-Assisted;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/10.341836
  • Filename
    341836