  • Abstract
    A real-time mapping system is described that allows instant display of both endocardial and epicardial activation as monitored by an array of transmural plunge-needle electrodes in the intact in situ heart. From the total cardiac muscle mass, 768 electrograms are recorded, with a programmable feature allowing the rapid and unrestricted selection of 110 endocardial electrograms for automated analysis. An activation analyzer (consisting of parallel analog circuitry) provides an activation image in real time, while video recorders provide both raw and processed data storage with the capability of immediate slow-motion replay. The nature of the data acquisition (110 plunge needles) limits this system to the experimental arrhythmia laboratory where its chief advantage is the enabling of the multichannel recording of endo data without the complications of open heart surgery. Equally important is the ability to automatically store the endocardial activation image locked in time to the simultaneously recorded epicardial image.
  • Keywords
    Circuits; Data acquisition; Displays; Electrodes; Heart; Image analysis; Instruments; Memory; Muscles; Real time systems; Arrhythmias, Cardiac; Computers; Electrocardiography; Endocardium; Heart; Humans;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.1987.326082
  • Filename
    4122570