• DocumentCode
    1219828
  • Title

    Does Physiological Blood Glucose Control Require an Adaptive Control Strategy?

  • Author

    Fischer, U. ; Schenk, W. ; Salzsieder, E. ; Albrecht, G. ; Abel, P. ; Freyse, E.-J.

  • Author_Institution
    Central Institute of Diabetes "Gerhardt Katsch,"
  • Issue
    8
  • fYear
    1987
  • Firstpage
    575
  • Lastpage
    582
  • Abstract
    To test the hypothesis that only an adaptive algorithm would guarantee optimal feedback control of glycemia in insulin-dependent diabetes, fasting chronically diabetic dogs at rest were subjected to short-term artificial beta cell treatment. Insulin was applied intravenously and an oral glucose load was given during the experiment. Employing the same dosing algorithm, three different control strategies were employed in a random order on different days: adaptive control (minimum variance controller, Test A), fixed command control using on-line parameter estimates (Test B), and fixed command control using off-line individually optimized dosage constants (Test C). Comparison was made to nondiabetic control animals. The glycemic profiles were entirely normal in Test A and C, but were distinctly elevated in Test B. The peripheral hyperinsulinaemia could, however, not be avoided by adaptive control. It is concluded that the restoration of physiological blood glucose control in insulin-dependent diabetes requires dosage parameters which are either continually adapted to the actual situation (adaptive control) or are optimized to meet the individual´s needs. In the latter case, fixed command control may be employed. Peripheral hyperinsulinaemia cannot be avoided as long as insulin is administered by a posthepatic route.
  • Keywords
    Adaptive algorithm; Adaptive control; Blood; Diabetes; Dogs; Feedback control; Insulin; Parameter estimation; Sugar; Testing; Animals; Biomedical Engineering; Blood Glucose; Diabetes Mellitus, Experimental; Dogs; Feedback; Female; Insulin Infusion Systems; Male;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.1987.326068
  • Filename
    4122594