DocumentCode
1219925
Title
An Open-Loop Computer-Based Drug Infusion System
Author
Martin, Roy W. ; Hill, Harlan F. ; Yee, Hsiao Ching ; Saeger, Louis C. ; Walter, M.Hildegard ; Chapman, C.Richard
Author_Institution
Department of Anesthesiology/Center for Bioengineering, University of Washington
Issue
8
fYear
1987
Firstpage
642
Lastpage
649
Abstract
This study reports the development of a computer-based infusion system and methodology to produce and maintain selected plasma concentrations. The method identifies pharmacokinetic infusion parameters for subjects from bolus injection response data, employs these values in control equations implemented by a portable microcomputer and computer-controlled infusion pump, and achieves and maintains selected stepwise drug levels by intravenous drug infusion. Infusion studies with four dogs and five humans resulted in correlation coefficients of 0.98 for the dogs and 0.94 for the humans, with rms errors in maintaining the drug concentration at the desired level of 13.4 and 19.3 percent, respectively. An analysis of error demonstrated that: 1) the control error was less than the value of the pharmacokinetic parameter estimation error in determining a single parameter value, 2) errors in several parameters can have cancelling or additive effects depending on their sign, and 3) an error in the sum of two of the model parameters (A and B) directly translates to equivalent time-independent error in the controlled level.
Keywords
Computer errors; Dogs; Drugs; Equations; Error correction; Humans; Microcomputers; Open loop systems; Plasmas; Portable computers; Alfentanil; Animals; Biomedical Engineering; Computers; Dogs; Fentanyl; Humans; Infusion Pumps;
fLanguage
English
Journal_Title
Biomedical Engineering, IEEE Transactions on
Publisher
ieee
ISSN
0018-9294
Type
jour
DOI
10.1109/TBME.1987.326097
Filename
4122603
Link To Document