• DocumentCode
    1219925
  • Title

    An Open-Loop Computer-Based Drug Infusion System

  • Author

    Martin, Roy W. ; Hill, Harlan F. ; Yee, Hsiao Ching ; Saeger, Louis C. ; Walter, M.Hildegard ; Chapman, C.Richard

  • Author_Institution
    Department of Anesthesiology/Center for Bioengineering, University of Washington
  • Issue
    8
  • fYear
    1987
  • Firstpage
    642
  • Lastpage
    649
  • Abstract
    This study reports the development of a computer-based infusion system and methodology to produce and maintain selected plasma concentrations. The method identifies pharmacokinetic infusion parameters for subjects from bolus injection response data, employs these values in control equations implemented by a portable microcomputer and computer-controlled infusion pump, and achieves and maintains selected stepwise drug levels by intravenous drug infusion. Infusion studies with four dogs and five humans resulted in correlation coefficients of 0.98 for the dogs and 0.94 for the humans, with rms errors in maintaining the drug concentration at the desired level of 13.4 and 19.3 percent, respectively. An analysis of error demonstrated that: 1) the control error was less than the value of the pharmacokinetic parameter estimation error in determining a single parameter value, 2) errors in several parameters can have cancelling or additive effects depending on their sign, and 3) an error in the sum of two of the model parameters (A and B) directly translates to equivalent time-independent error in the controlled level.
  • Keywords
    Computer errors; Dogs; Drugs; Equations; Error correction; Humans; Microcomputers; Open loop systems; Plasmas; Portable computers; Alfentanil; Animals; Biomedical Engineering; Computers; Dogs; Fentanyl; Humans; Infusion Pumps;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.1987.326097
  • Filename
    4122603