Title :
Hybrid stimulator for chronic experiments
Author :
Del Pozo, F. ; Delgado, J.M.R.
Author_Institution :
Center "Ramon y Cajal", Madrid, Spain
Abstract :
A hybrid stimulator is described which combines constant current and constant voltage characteristics. Pulses are delivered with high output impedance to assure the passage of a predetermined amount of current. During the quiescent periods between pulses, the output changes to low impedance, diminishing the post-pulse charge. Electrode erosion and tissue damage are thus minimized, permitting safe, long term excitation.
Keywords :
Corrosion; Electric resistance; Electrodes; FETs; Impedance; Instruments; Pulse measurements; Steel; Switches; Voltage; Animals; Computers; Electrodes; Humans; Neurons;
Journal_Title :
Biomedical Engineering, IEEE Transactions on
DOI :
10.1109/TBME.1978.326384