DocumentCode :
1221974
Title :
A Method for Measuring Functional Residual Capacity in Neonates with Endotracheal Tubes
Author :
Schwartz, Jacob G. ; Fox, William W. ; Shaffer, Thomas H.
Author_Institution :
Department of Biomedical Engineering and Science, Drexel University
Issue :
3
fYear :
1978
fDate :
5/1/1978 12:00:00 AM
Firstpage :
304
Lastpage :
307
Abstract :
This study evaluates a new 60-s closed circuit helium (He) dilution technique for determination of functional residual capacity (FRC) in intubated neonates independent of snall gas leaks present around uncuffed endotracheal (ET) tubes. By analyticaly relating the fall in He concentration due to mixing with that due to leakage it is possible to predict the final equflibration concentration of He and, therefore, correct for ET tube leaks. The system (120 ml) contains an air pump, He meter, breathing bag in cylinder, a strip chart readout, and solenoid value. Continuous positive airway pressure (CPAP) or ventilator pressure can be applied during testing. One hundred in vitro measurements of FRC ranging from 5-100 cc in both leak and nonleak models were performed and were accurate to within ±7.8% standard deviation. Functional residual capacity measurements were also performed in 30 infants (weight 600-4400 gm).
Keywords :
Biomedical measurements; Circuits; Helium; In vitro; Jacobian matrices; Lungs; Medical diagnostic imaging; Pediatrics; Performance evaluation; Volume measurement; Evaluation Studies as Topic; Functional Residual Capacity; Helium; Humans; Infant, Newborn; Lung Volume Measurements; Methods; Respiratory Distress Syndrome, Newborn; Trachea;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/TBME.1978.326310
Filename :
4122835
Link To Document :
بازگشت