• DocumentCode
    1222708
  • Title

    C-Mode Scan and Resolution Improvement Techniques for Ultrasonic Diagnosis

  • Author

    Ito, Ken-Ichi ; Itoh, Masayasu ; Yuta, Shin-Ichi ; Yokoi, Hiromu ; Inouye, Tamon

  • Author_Institution
    Department of Electronic Engineering, Tokyo University of Agriculture and Technology
  • Issue
    1
  • fYear
    1979
  • Firstpage
    11
  • Lastpage
    17
  • Abstract
    This paper presents a new method for obtaining plane images perpendicular to the axis of the transducer and discusses some resolution improvement techniques utilizing a computer. The method, called C-mode scan, is carried out by successive application of a conventional B-mode scan, and supplies more information such as the outline, size, and extent of a lesion such as a breast tumor. To improve the resolution of a transducer, a deconvolution technique together with non-linear filtering are employed from a spectral point of view. Several clinical images produced by C-mode and the computer improvement of resolution are demonstrated.
  • Keywords
    Deconvolution; Displays; Dynamic range; Image processing; Image resolution; Image storage; Indium tin oxide; Monitoring; Tomography; Ultrasonic transducers; Action Potentials; Algorithms; Electric Stimulation; Electromyography; Evoked Potentials, Motor; Humans; Male; Middle Aged; Models, Neurological; Models, Statistical; Motor Neurons; Reproducibility of Results; Sensitivity and Specificity; Signal Processing, Computer-Assisted;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.1979.326490
  • Filename
    4122922