• DocumentCode
    1224437
  • Title

    Transthoracic Ventricular Defibrillation in the 100 kg Calf with Untruncated and Truncated Exponential Stimuli

  • Author

    Schuder, John C. ; Gold, Jerry H. ; Stoeckle, Harry ; Granberg, Thomas A. ; Dettmer, Joseph C. ; Larwill, Marcus H.

  • Author_Institution
    Department of Surgery, University of Missouri
  • Issue
    1
  • fYear
    1980
  • Firstpage
    37
  • Lastpage
    43
  • Abstract
    The effectiveness in reversing ventricular fibrillation of 30 s duration of 50 untruncated and truncated exponential waveforms was studied in 6000 fibrillation-defibrillation episodes in 100 kg calves. The study employed waveforms which could be generated with nominal stored energy levels (in the basic capacitor-switch sense) of 220, 440, and 660 J and initial current levels of 50, 70, 100, 140, and 200 A. Fifty A waveforms having a time constant of decay of 10 ms were truncated at 15 and 30 A while those having time constants of 20 and 30 ms were truncated at 15, 30, and 45 A. Seventy A waveforms having time constants of 5, 10, and 15 ms were untruncated or truncated at 15, 30, and 45 A. One hundred A waveforms having time constants of 2.5, 5.0, and 7.5 ms were untruncated or truncated at 15, 30, and 45 A. One hundred forty A waveforms having time constants of 1.25, 2.50, and 3.75 ms were untruncated or truncated at 15, 30, and 45 A. Two hundred A waveforms having time constants of 0.63, 1.25, and 1.88 ms were untruncated or truncated at 45 A. We used 75 animals in the study.
  • Keywords
    Animals; Capacitors; Defibrillation; Dogs; Electric shock; Energy states; Fibrillation; Gold; Pulse shaping methods; Switches; Animals; Cattle; Electric Countershock;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.1980.326689
  • Filename
    4123124