Title :
An AC bridge readout for bolometric detectors
Author :
Rieke, F.M. ; Lange, A.E. ; Beeman, J.W. ; Haller, E.E.
Author_Institution :
Dept. of Phys., California Univ., Berkeley, CA, USA
fDate :
2/1/1989 12:00:00 AM
Abstract :
The authors have developed a bolometer readout circuit which greatly improves the low-frequency stability of bolometric detectors. The circuit uses an AC bias voltage and two matched bolometers and allows stable DC bolometer operation for integration times ⩾10 s. In astronomical applications the readout allows for qualitatively different observation modes (e.g. staring or slow-drift scanning) which are particularly well suited for space observations and for the use of arrays. In many applications the readout can increase sensitivity. The authors present noise spectra for 4He temperature bolometers with no excess noise at frequencies greater than 0.1 Hz. The measured optical responsivity of a bolometer operated with the present readout is the same as that of a bolometer operated with a conventional readout
Keywords :
bolometers; bridge circuits; AC bridge readout; astronomical applications; bolometric detectors; low-frequency stability; noise spectra; stable DC bolometer operation; Bolometers; Bridge circuits; Circuit stability; Detectors; Extraterrestrial measurements; Frequency; Helium; Optical noise; Temperature sensors; Voltage;
Journal_Title :
Nuclear Science, IEEE Transactions on