• DocumentCode
    1225452
  • Title

    Abdominal-Lead Fetal Electrocardiographic R-Wave Enhancement for Heart Rate Determination

  • Author

    Azevedo, Stephen ; Longini, Richard L.

  • Author_Institution
    Biomedical Engineering Program, Carnegie-Mellon University
  • Issue
    5
  • fYear
    1980
  • fDate
    5/1/1980 12:00:00 AM
  • Firstpage
    255
  • Lastpage
    260
  • Abstract
    The reliability of fetal heart rate (FHR) from external abdominal-lead electrocardiographic waveforms depends upon the R-wave detection capability of the monitoring device. Assuming that the maternal component of the ECG is canceiled by techniques described elsewhere [1]-[4], we are left with a fetal electrocardiogm (FECG) which may be conrupted by high noise levels. This paper describes several digital signal processing techniques by which the detection of fetal R-waves is improved. These techniques-narrow band-pass filtering, local peak differencing, autoregressive filtering, and matched filtering-are theoretically developed. The performance of each technique is measured and analyzed for two sample fetal ECG inputs.
  • Keywords
    Abdomen; Band pass filters; Electrocardiography; Fetal heart rate; Filtering; Heart rate; Heart rate detection; Heart rate measurement; Matched filters; Noise level; Analog-Digital Conversion; Electrocardiography; Fetal Heart; Fetal Monitoring; Forecasting; Heart Rate; Humans; Image Enhancement;