Quantitation of PET data with the EM reconstruction technique
Author :
Rosenqvist, Goran ; Dahlbom, Magnus ; Eriksson, Lars ; Bohm, Christian ; Blomqvist, Gunnar
Author_Institution :
Karolinska Hospital, Stockholm, Sweden
Volume :
36
Issue :
1
fYear :
1989
fDate :
2/1/1989 12:00:00 AM
Firstpage :
1113
Lastpage :
1116
Abstract :
The capability of a modified EM (expectation maximization) algorithm to quantify image data has been tested in phantom studies and in two clinical cases, cerebral blood flow studies and dopamine D2-receptor studies. Data from phantom studies indicate the superiority of images reconstructed with the EM technique compared to images reconstructed with the conventional filtered back-projection (FB) technique in areas with low statistics. At higher statistics the noise characteristics to the two technique coincide. Clinical data support these findings. Functional image data were generated regarding cerebral blood flow and receptor/dissociation constant (Bmax/ kD). The functional data based on EM reconstructed images show generally less noise and higher spatial resolution than functional data generated by FB reconstructed images
Keywords :
computerised tomography; radioisotope scanning and imaging; PET; cerebral blood flow studies; dopamine D2-receptor studies; expectation maximization; filtered back-projection; image data; phantom studies; positron emission tomography; Acceleration; Blood flow; Convergence; Event detection; Image reconstruction; Imaging phantoms; Noise level; Positron emission tomography; Spatial resolution; Statistics;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/23.34614
Filename :
34614
Link To Document :
بازگشت