DocumentCode
1226002
Title
Experimental evaluation of a modular scintillation camera for SPECT
Author
Rogers, W.L. ; Clinthorne, N.H. ; Shao, L. ; Koral, K.F.
Author_Institution
Div. of Nucl. Med., Michigan Univ. Med. Sch., Ann Arbor, MI, USA
Volume
36
Issue
1
fYear
1989
fDate
2/1/1989 12:00:00 AM
Firstpage
1122
Lastpage
1126
Abstract
A modular scintillation camera has been constructed for use in a ring-type single photon tomograph for nuclear medicine. Its performance has been characterized, and experiments with coupling the x and y maximum-likelihood position estimators and varying the sampling of the scintillation light are described. Average spectral resolution in the x direction is equivalent to a 3.6-mm FWHM (full width at half-maximum) Gaussian. In the y direction the response function averages 3.6 mm FWHM. RMS (root mean square) linearity is 0.9 mm in x and 1.1 mm in y . Potential for improving resolution by increasing the sampling density in x has been demonstrated, but a two-dimensional position estimator would be required to maintain acceptable linearity. It is concluded that with the exception of reduced resolution within 6-9 mm of the edge, the camera performance is within design specifications
Keywords
biomedical equipment; computerised tomography; iterative methods; radioisotope scanning and imaging; maximum-likelihood position estimators; modular scintillation camera; nuclear medicine; response function; ring-type single photon tomograph; sampling density; scintillation light; spectral resolution; Bars; Cameras; Geometry; Linearity; Maximum likelihood detection; Noise shaping; Optical coupling; Photoelectricity; Sampling methods; Single photon emission computed tomography;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/23.34616
Filename
34616
Link To Document