• DocumentCode
    1226002
  • Title

    Experimental evaluation of a modular scintillation camera for SPECT

  • Author

    Rogers, W.L. ; Clinthorne, N.H. ; Shao, L. ; Koral, K.F.

  • Author_Institution
    Div. of Nucl. Med., Michigan Univ. Med. Sch., Ann Arbor, MI, USA
  • Volume
    36
  • Issue
    1
  • fYear
    1989
  • fDate
    2/1/1989 12:00:00 AM
  • Firstpage
    1122
  • Lastpage
    1126
  • Abstract
    A modular scintillation camera has been constructed for use in a ring-type single photon tomograph for nuclear medicine. Its performance has been characterized, and experiments with coupling the x and y maximum-likelihood position estimators and varying the sampling of the scintillation light are described. Average spectral resolution in the x direction is equivalent to a 3.6-mm FWHM (full width at half-maximum) Gaussian. In the y direction the response function averages 3.6 mm FWHM. RMS (root mean square) linearity is 0.9 mm in x and 1.1 mm in y. Potential for improving resolution by increasing the sampling density in x has been demonstrated, but a two-dimensional position estimator would be required to maintain acceptable linearity. It is concluded that with the exception of reduced resolution within 6-9 mm of the edge, the camera performance is within design specifications
  • Keywords
    biomedical equipment; computerised tomography; iterative methods; radioisotope scanning and imaging; maximum-likelihood position estimators; modular scintillation camera; nuclear medicine; response function; ring-type single photon tomograph; sampling density; scintillation light; spectral resolution; Bars; Cameras; Geometry; Linearity; Maximum likelihood detection; Noise shaping; Optical coupling; Photoelectricity; Sampling methods; Single photon emission computed tomography;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.34616
  • Filename
    34616